Abstract
Abstract
Objective
Test performance of a new prototype microdebrider blade. The commercially‐available, standard 4.0‐mm Medtronic straight Tricut blade was tested against the new, prototype 4.3‐mm Medtronic straight Quadcut blade in experimental surgical conditions utilizing both a nasal polyp (NP) analog (raw oysters) and an allergic fungal sinusitis (AFS) analog (minced beef cat food).
Methods
Both sides of 5 thawed, fresh‐frozen cadaver heads were utilized for NP analog removal. AFS analog was removed out of specimen containers. A total of 10 paired data points were collected for time of surgical removal and number of clogs comparing the 2 blades.
Results
For simulated AFS debris, the prototype clogged less than the standard blade (0 vs 4.5, p p p Conclusion
The prototype blade offers faster debris and polyp removal and clogs less in the setting of allergic fungal debris. As the new blade is faster and more aggressive, introduction into a surgical setting should be guarded with an appropriate degree of caution. © 2011 ARS‐AAOA, LLC.